баннер
Дом блог

EC-105HTP,MTP,MTHP, High and low temperature constant temperature bath (1000L)

EC-105HTP,MTP,MTHP, High and low temperature constant temperature bath (1000L)

November 14, 2014

EC-105HTP,MTP,MTHP, High and low temperature constant temperature bath (1000L)

Project

Type

Series

HT

MT

MTH

function

Temperature occurs in a way

Dry wet bulb method

Temperature range

-20 ~ + 100 ℃

-40 ~ + 100 ℃

-40 ~ + 150 ℃

Range of temperature 

Below the + 100℃

± 0.3 ℃

+Above the 101℃

± 0.5 ℃

Temperature distribution

Below the + 100℃

± 1.0 ℃

Above the + 101℃

± 2.0 ℃

The temperature drops the time

+20 ~ -20 ℃

Within 60 minutes

+20 ~ -40 ℃

Within 9 0 minutes

+20 ~ -40 ℃

Within 9 0 minutes

Temperature rise time

-20 ~ + 100 ℃

Within 45 minutes

-40 ~ + 100 ℃

Within 50 minutes

-40 ~ + 150 ℃

Within 75 minutes

The internal volume of the uterus was tested

1000L

Test room inch method (width, depth and height)

1000mm × 1000mm × 1000mm

Product inch method (width, depth and height)

1400mm × 1370mm × 1795mm

Make the material

External outfit

Test room control panel

machine room

Cold steel plate, cold steel plate beige

(Color table 2.5Y8 / 2)

Inside

Stainless steel plate (SUS304,2B polished)

Broken heat material

Test room

Hard synthetic resin

glass wool

door

Hard synthetic resin foam cotton, glass cotton

Project

Type

Series

HT

MT

MTH

Cooling dehumidifying device

 Cooling-down method

Mechanical section shrinkage mode

 Cooling medium

R404A

compressor

Output (number of units)

0.75kW (1)

1.5kW (1)

Cooling and dehumidifier

Multi-channel mixed heat sink type

The condenser

Multi-channel mixed radiator plate type (air cooling type)

Calorifier

Form

Nickel-chromium heat-resistant alloy heater

Volume

3.5kW

Blower

Form

Multi-channel mixed radiator plate type (air cooling type)

Motor capacity

40W

 Controller

The temperature is set

-22.0 ~ + 102.0 ℃

-42.0 ~ + 102.0 ℃

-42.0 ~ + 152.0 ℃

Humidity is set

0 ~ 98%RH (But the temperature of the wet and dry bulb is 10-85 ℃ )

Time setting Fanny

0 ~ 999 Time 59 minutes (formula) 0 ~ 20000 Time 59 minutes (formula formula)

Set decomposition energy

Temperature 0.1℃, humidity 1% RH for 1 min

Indicate accuracy

Temperature ± 0.8℃ (tp.), humidity ± 1% RH (tp.), time ± 100 PPM

Vacation type

Value or program

Stage number

20-stage / 1 program

The number of procedures

The maximum number of incoming force (RAM) programs is 32 programs

The maximum number of internal ROM programs is 13 programs

Round-trip number

 98 times maximum or unlimited

Number of round-trip repeats

Maximum 3 times

Displace the end

Pt 100Ω(at 0 ℃), gradeB(JIS C 1604-1997)

Control action

When splitting the PID action

Endovirus function

Early delivery function, standby function, setting value maintenance function, power outage protection function,

Power action selection function, maintenance function, transportation round-trip function,

Time delivery function, time signal output function, overrising and overcooling prevention function,

Abnormal representation function, external alarm output function, setting paradigm representation function,

Transport type selection function, the calculation time represents the function, the slot lamp lamp function

Project

Type

Series

HT

MT

MTH

Control panel

Equipment machine

LCD operating panel (type contact panel),

Represents lamp (power, transport, abnormal), test power supply terminal, external alarm terminal,

Time signal output terminal, power cord connector

 Protective device

Refrigerating cycle

Overload protection device, high blocking device

Calorifier

Temperature over-rise protection device, temperature fuse

Blower

Overload protection device

Control panel

Leakage breaker for power supply, fuse (for heater, humidifier),

Fuse (for operating loop), temperature rise protection device (for testing),

Temperature rise overcooling prevention device (test material, in microcomputer)

Offproducts (sets)

House receiving (4), house board (2), operation instructions (1)

Equipment products

Adventitia

hard borosilicate glass  270mm× 190mm

 

Cable hole

内径  50mm

 

The trough inside the lamp

AC100V 15W White hot ball

 

Wheel

 

 

Horizontal adjustment

 

 

Electrovirus characteristics 

Source *

 AC  three-phase 380V  50Hz

Maximum load current

13A

15A

Capacity of the leakage breaker for the power supply

25 A

Sensory current 30mA

Power distribution thickness

8mm2

14mm2

Rubber insulation hose

Coarseness of grounding wire

3.5mm2

5.5mm2

 Tubing

drain-pipe *

PT1/2

Product weight

470kg

540kg